Page 1: Safety Information Instruction Sheet Introduction The Fluke Loop Calibrator is a source and measurement tool for 0 to 20 mA or 4 to 20 mA current loop testing and a measurement tool for dc voltage from 0 to 28 V.

The calibrator does not source and measure simultaneously. Page 2: Explanation Of International Symbols Warning To avoid possible electric shock or personal injury: Never apply more than 30 V between any two terminals, or between any terminal and earth ground.

Inspect the test leads for damaged insulation or exposed metal. Check lead continuity. Replace damaged test leads before you use the meter. Press while turning on the calibrator to select loop power. The Loop symbol appears in the display. Press simultaneously to start the SpanCheck function. The symbol appears in the display.

Page 4 Sourcing mA Use source mode whenever you need to supply current into a passive circuit. Insert the test leads into the terminals as shown below. Note A path must exist for current to flow between the output and mA terminals, otherwise the display will indicate an overload OL when you set an output value.

This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse or abnormal conditions of operation or handling. Page 6: Simulating A Transmitter Simulating a Transmitter When simulating the operation of a transmitter, the calibrator regulates the loop current to a known value that the user selects.

A 12 to 28 V loop supply must be available. Insert the test leads as shown below. To exit, turn the calibrator off or to V. Page Measuring Dc Volts Do not allow water into the case. Remove any input signals prior to removing test leads and opening case. When servicing the calibrator, use only specified replacement parts. For maintenance procedures not described in this sheet, contact a Fluke Service Center.

Replace as needed. If the calibrator needs repair, contact a Fluke Service Center. If the calibrator is under warranty, see the warranty statement for terms. If the warranty has lapsed, the calibrator will be repaired and returned for a fixed fee.